• Skip to primary navigation
  • Skip to main content
  • Skip to primary sidebar
  • Skip to footer

MassDevice

The Medical Device Business Journal — Medical Device News & Articles | MassDevice

  • Special Reports
  • Technologies
    • Artificial Intelligence (AI)
    • Cardiovascular
    • Orthopedics
    • Neurological
    • Diabetes
    • Surgical Robotics
  • Business & Finance
    • Wall Street Beat
    • Earnings Reports
    • Funding Roundup
    • Mergers & Acquisitions
    • Initial Public Offering (IPO)
    • Legal News
    • Personnel Moves
    • Medtech 100 Stock Index
  • Regulatory
    • Food & Drug Administration (FDA)
    • Recalls
    • 510(k)
    • Pre-Market Approval (PMA)
    • MDSAP
    • Clinical Trials
  • Resources
    • About MassDevice
    • Leadership in Medtech
    • Manufacturers & Suppliers Search
    • MedTech100 Index
    • Videos
    • Webinars
    • Whitepapers
    • Voices
    • In-Depth Coverage
    • Latest News
  • Attend DeviceTalks
    • Events
      • DeviceTalks Minnesota – May 4
      • DeviceTalks Boston – May 27–28
      • DeviceTalks West
      • DeviceTalks Tuesdays
    • DeviceTalks Podcast Network
      • DeviceTalks Weekly
      • AbbottTalks
      • Boston ScientificTalks
      • DeviceTalks AI
      • IntuitiveTalks
      • MedtechWOMEN Talks
      • MedtronicTalks
      • Neuro Innovation Talks
      • Ortho Innovation Talks
      • Structural Heart Talks
      • StrykerTalks
  • Advertise
  • Subscribe
Home » Teleflex has a serious balloon catheter recall, 3 deaths linked to issue

Teleflex has a serious balloon catheter recall, 3 deaths linked to issue

June 14, 2024 By Sean Whooley

Teleflex Arrow FiberOptix intra-aortic balloon IAB catheter kit
The Arrow FiberOptix intra-aortic balloon catheter. [Image from the Teleflex website]
The FDA says a recall of Teleflex (NYSE:TFX) intra-aortic balloon (IAB) catheter kits is Class I, the most serious kind of recall.

Teleflex recalled its Arrow FiberOptix IAB catheter kit and Arrow UltraFlex IAB catheter kit. This recall is a correction, not a product removal. However, the issues related to the kits resulted in reports of 31 injuries and three deaths, the FDA said in a notice.

Arrow IABs work with a balloon pump for patients undergoing cardiac and non-cardiac surgery. They also treat adult patients who have acute coronary syndrome or complications of heart failure. According to the FDA, Teleflex recalled these specific pumps after finding a manufacturing error that may cause the balloon to become overtwisted.

Users can sometimes identify overtwisting visually, but the issue could affect a device even if the user fails to see it. The issue may prevent the balloon from fully inflating, cause blood to back up in the tubing and allow helium to leak. This may lead to catheter damage or insertion difficulty during use.

The issue may cause serious injury, including blood loss, tearing in the artery (perforation), unstable blood pressure (hemodynamic instability), prevention of blood flow to the heart (myocardial ischemia), or death. Teleflex received 322 complaints that included the 31 injuries and three deaths related to the issue.

Teleflex initiated the recall on April 29, 2024, recalling 16,959 devices distributed in the U.S. between May 7, 2022, and April 8, 2024.

Recommendations to Teleflex customers regarding this recall

In an urgent medical device notification, Teleflex instructed customers to ensure the availability of a back-up IAB catheter before using the device. They should also inspect all IAB catheters for signs of overtwisted balloon wraps or bent balloon shafts.

Additionally, customers should use fluoroscopic guidance to insert all IAB catheters in the scope of the recall. They must use fluoroscopy to assess the completeness of balloon inflation, once inserted and connected.

Teleflex also advised vigilance in responding to pump alarms and bedside indicators that could mean unexpected performance. That could include blood in the helium pathway and lower-than-expected diastolic pressure augmentation.

If customers identify the issue, they should remove the catheter and replace it as outlined in the instructions for use. They may insert a replacement contralaterally or ipsilaterally after considering individual femoral access options and relevant clinical features. Customers should also report every suspected incidence of the condition to Teleflex as soon as practical.

Filed Under: Cardiac Implants, Cardiovascular, Catheters, Featured, Food & Drug Administration (FDA), Recalls, Regulatory/Compliance, Structural Heart Tagged With: FDA, Teleflex

More recent news

  • Butterfly Network to chase ultrasound brain-computer interfaces with newcomer Bridge Neurotech
  • Apple must pay an extra $184M to Masimo, judge says
  • MMI grasps $125M to support robotic microsurgery system
  • FDA grants de novo nod to handheld ER cardiac implant identifier
  • J&J delivers Shockwave data sustaining below-the-knee gains for two years

About Sean Whooley

Sean Whooley is a senior editor who mainly produces work for MassDevice, Medical Design & Outsourcing and Drug Delivery Business News. He received a bachelor's degree in multiplatform journalism from the University of Maryland, College Park. You can connect with him on LinkedIn or email him at [email protected].

Primary Sidebar

“md
EXPAND YOUR KNOWLEDGE AND STAY CONNECTED
Get the latest med device regulatory, business and technology news.

DeviceTalks Weekly

See More >
MDO ad
MDO ad

Footer

MASSDEVICE MEDICAL NETWORK

DeviceTalks
Drug Delivery Business News
Medical Design & Outsourcing
Medical Tubing + Extrusion
Drug Discovery & Development
Pharmaceutical Processing World
MedTech 100 Index
R&D World
Medical Design Sourcing

DeviceTalks Webinars, Podcasts, & Discussions

Attend our Monthly Webinars
Listen to our Weekly Podcasts
Join our DeviceTalks Tuesdays Discussion

MASSDEVICE

Subscribe to MassDevice E-Newsletter
Advertise with us
About
Contact us

Copyright © 2026 · Arrowfly LLC and its licensors. All rights reserved.
The material on this site may not be reproduced, distributed, transmitted, cached or otherwise used, except with the prior written permission of Arrowfly.

Privacy Policy