Zoll reminds on AED battery issue

December 22, 2009 by MassDevice staff

Zoll Medical Corp. issues a reminder about a potential problem with the batteries in some of its automated external defibrillators, which may have played a role in the deaths of two people.

Zoll logo

Zoll Medical Corp. (NSDQ:ZOLL) is reminding customers about defective batteries and a software glitch in one of its defibrillators that may have killed two people.

In April, the Chelmsford, Mass.-based resuscitation device maker issued a "voluntary worldwide field corrective action" on its AED Plus automated external defibrillator after discovering the problem.

Batteries in some of the devices made before Feb. 12 are defective and the software in the debrillators that's supposed to detect such problems doesn't work properly, the company said at the time. The roughly 180,000 affected devices carry serial numbers below "X_ _ _200000," according to the warning, and defibrillators older than three years — about 80,000 of them — are the most likely to fail.

On Dec. 21 Zoll issued a reminder about the problem, saying customers who haven't run a software upgrade and notified the company must do so immediately.

The problem initially came to light from customer reports, according to a press release, and "there has been one clinical event reported in which a defibrillation shock was not delivered [and] the patient subsequently died."

A subsequent review found that three other devices failed and, in one of those instances, a second patient later died.

The software defect can be fixed with a patch, available for download from the company's website, that can detect defective batteries and alert users.

Zoll said customers should take the following actions:

  1. All AED Plus devices with serial numbers below “X_ _ _200000” must be upgraded with the latest software release. The software update can be easily downloaded from the company’s web site and installed by the user.
  2. Users must send Zoll an acknowledgement that the corrective action has been implemented after updating the device software. This confirmation is required for customers to avoid liability and for Zoll to ensure defective devices have been remedied.

Comments

Comment viewing options

Select your preferred way to display the comments and click "Save settings" to activate your changes.
browse jobs and career opportunities in the medical devices industry
view counter
Free Literature and Charts from Emergo Group
view counter

NEWS

Medtronic vet guides Cameron Health

Warren Watson, a 30-year Medtronic Inc. employee with family connections to founder Earl Bakken, is...

Quidel gains FDA clearance for pregnancy test

Quidel Corp. wins Food & Drug Administration 510(k) clearance for its chorionic gonadotropin-...

Ex-FDA commissioner joins board of Michigan cancer treatment firm

Ann Arbor, Mich.-based HistoSonics Inc. snags former Food & Drug Administration commissioner Dr....

FDA releases July 510(k) clearances

The Food & Drug Administration releases a list of the medical devices it granted 510(k) clearance...

Feds indict former Spectranetics execs

A federal grand jury indicts former Spectranetics CEO John Schulte, two other executives and a...

SEARCH JOBS


Search our database for the best open jobs in the medical devices industry.

keywords

job title or skills
location

city, state or zip

Jobs by MassDevice

view counter
view counter
Instron manufactures tensile, compression, fatigue, impact and hardness testing machines, testing a variety of  materials and ra
view counter
Built on an AdaptiveTheme using Drupal by Michael Knapp  mknapp