FDA, Edwards issue Class I recall of hemodialysis system

January 29, 2010 by MassDevice staff

The Food & Drug Administration and Edwards Lifesciences issue a Class I recall, the highest-level callback, for Edwards's Aquarius hemodialysis system distributed by Baxter International.

FDA logo

The Food & Drug Administration and Edwards Lifesciences Inc. (NYSE:EW) issued a Class I recall of the Aquarius hemodialysis system after reports came in of "clinically significant fluid imbalance."

The device, used to remove waste products and extra fluid from the blood of patients with kidney failure, is designed to monitor the flow of fluid in and out of patients' bodies and to trigger an alarm if it detects an imbalance. Patients can override the alarm without fixing the problem that caused it (a kink in the line, for example), causing an increase or decrease in patients' blood volume.

Irvine, Calif.-based Edwards sent a letter to customers Jan. 11 advising them of the potential problem and the steps to take to correct it, including its plans to release a software upgrade to prevent patients from overriding the alarm more than five times within a 20-minute period.

A Class I recall is the highest level of callback and means the federal watchdog agency believes there's a possibility the recalled device could cause serious injury or death. Edwards spokeswoman Amanda Fowler told MassDevice that no adverse effects were reported as a result of the potential problem as of March 2009 and said the company volunteered to upgrade the software then. That upgrade won 510(k) clearance from the FDA last November, Fowler said, but hasn't yet been put in place.

The recall includes the following model numbers, distributed between July 12, 2007 and March 18, 2009, by Baxter International Inc. (NYSE:BAX): GEF08200, GEF09500, GEF09600, GEF09700, and GEF09800, using Software version 6.00.04.

Support the news, become a MassDevice.com insider
view counter
META's Needs Assessment System measures the gap between ideal and current sales performance
view counter

NEWS

Medtronic vet guides Cameron Health

Warren Watson, a 30-year Medtronic Inc. employee with family connections to founder Earl Bakken, is...

Quidel gains FDA clearance for pregnancy test

Quidel Corp. wins Food & Drug Administration 510(k) clearance for its chorionic gonadotropin-...

Ex-FDA commissioner joins board of Michigan cancer treatment firm

Ann Arbor, Mich.-based HistoSonics Inc. snags former Food & Drug Administration commissioner Dr....

FDA releases July 510(k) clearances

The Food & Drug Administration releases a list of the medical devices it granted 510(k) clearance...

Feds indict former Spectranetics execs

A federal grand jury indicts former Spectranetics CEO John Schulte, two other executives and a...

SEARCH JOBS


Search our database for the best open jobs in the medical devices industry.

keywords

job title or skills
location

city, state or zip

Jobs by MassDevice

view counter
view counter
Free Literature and Charts from Emergo Group
view counter
Built on an AdaptiveTheme using Drupal by Michael Knapp  mknapp